Baxter / SIGMA Spectrum Infusion System PUBLISHED

Description (CNA)

An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Affected (CNA)

Vendor / productVersions
Baxter / SIGMA Spectrum Infusion Systemaffected 6.05 (model 35700BAX) with wireless battery module (WBM) version 16

CWE