← CVE Intelligence

CVE-2015-3952

Hospira / Plum A+ Infusion System PUBLISHED

Description (CNA)

Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Affected (CNA)

Vendor / productVersions
Hospira / Plum A+ Infusion Systemaffected <= 13.4
Hospira / Plum A+3 Infusion Systemaffected <= 13.6
Hospira / Symbiq Infusion Systemaffected <= 3.13

CWE

CWE-312

References

ADP enrichment

CVE

No English description supplied.