← CVE Intelligence

CVE-2018-10631

Medtronic N'Vision Clinician Programmer Protection Mechanism Failure PUBLISHED

Description (CNA)

The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.

Affected (CNA)

Vendor / productVersions
Medtronic / 8840 N’Vision Clinician Programmeraffected All versions
Medtronic / 8870 N’Vision removable Application Cardaffected All versions

CWE

CWE-693

References

ADP enrichment

CVE

No English description supplied.

CISA-ADP

No English description supplied.

No references supplied.