CVE-2019-10964

Medtronic MiniMed 508 and Paradigm Series Insulin Pumps Improper Access Control PUBLISHED

Description (CNA)

Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

Affected (CNA)

Vendor / productVersions
Medtronic / MiniMed 508 pumpaffected All versions
Medtronic / MiniMed Paradigm 511 pumpaffected All versions
Medtronic / MiniMed Paradigm 512/712 pumpsaffected All versions
Medtronic / MiniMed Paradigm 712E pumpaffected All versions
Medtronic / MiniMed Paradigm 515/715 pumpsaffected All versions
Medtronic / MiniMed Paradigm 522/722 pumpsaffected All versions
Medtronic / MiniMed Paradigm 522K/722K pumpsaffected All versions
Medtronic / MiniMed Paradigm 523/723 pumpsaffected 0 through Software Versions 2.4A
Medtronic / MiniMed Paradigm 523K/723K pumpsaffected 0 through Software Versions 2.4A
Medtronic / MiniMed Paradigm Veo 554/754 pumpsaffected 0 through Software Versions 2.6A
Medtronic / MiniMed Paradigm Veo 554CM/754CM pumpsaffected 0 through Software Versions 2.7A

CWE

CWE-284