← CVE Intelligence

CVE-2019-13531

Medtronic Valleylab FT10 and LS10 Improper Authentication PUBLISHED

Description (CNA)

In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.

Affected (CNA)

Vendor / productVersions
Medtronic / Valleylab FT10 Energy Platform (VLFT10GEN)affected 0 through 2.1.0; affected 0 through 2.0.3
Medtronic / Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States)affected 0 through 1.20.2

CWE

CWE-287

References

ADP enrichment

CVE

No English description supplied.