Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface PUBLISHED

Description (CNA)

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

Affected (CNA)

Vendor / productVersions
Dräger / Perseus A500affected 2.00; affected 2.01; affected 2.02; unaffected 2.03

CWE

CWE-1286