← CVE Intelligence

CVE-2020-27252

Medtronic MyCareLink Smart Time-of-check Time-of-use Race Condition PUBLISHED

Description (CNA)

Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.

Affected (CNA)

Vendor / productVersions
Medtronic / Smart Model 25000 Patient Readeraffected All versions

CWE

CWE-367

References

ADP enrichment

CVE

No English description supplied.