Description (CNA)
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.
Affected (CNA)
| Vendor / product | Versions |
|---|---|
| Dräger / Infinity Acute Care System | affected 0 before VG4.2; affected 0 before VG4.1.1; affected 0 before VG4.0.3 |
| Dräger / Standalone Infinity M540 patient monitor | affected 0 before VG4.2; affected 0 before VG4.1.1 |
CWE
CWE-345
References
- https://static.draeger.com/security/download/PSA-19-255-03_Product-Security-Advisory-IACS.pdf
- https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofed-network-message-handling-dos-tampering
ADP enrichment
No English description supplied.
No references supplied.