Description (CNA)
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Affected (CNA)
| Vendor / product | Versions |
|---|---|
| Baxter / Life2000 Ventilation System | affected 06.08.00.00 and prior |