Description (CNA)
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Affected (CNA)
| Vendor / product | Versions |
|---|---|
| Radiometer Medical Aps / ABL90 FLEX and ABL90 FLEX PLUS Analyzers | affected Windows 7 Operating system; affected Windows XP Operating system |
| Radiometer Medical Aps / AQT90 FLEX Analyzers | affected Windows 7 Operating system; affected Windows XP Operating system |
| Radiometer Medical Aps / ABL800 BASIC and ABL800 FLEX analyzers | affected Windows 7 Operating system; affected Windows XP Operating system |
CWE
CWE-798, CWE-250
References
ADP enrichment
No English description supplied.
No references supplied.