← CVE Intelligence

CVE-2025-24836

Qardio Heart Health IOS and Android Application and QardioARM A100 Uncaught Exception PUBLISHED

Description (CNA)

With a specially crafted Python script, an attacker could send continuous startMeasurement commands over an unencrypted Bluetooth connection to the affected device. This would prevent the device from connecting to a clinician's app to take patient readings and ostensibly flood it with requests, resulting in a denial-of-service condition.

Affected (CNA)

Vendor / productVersions
Qardio / Heart Health IOS Mobile Applicationaffected 2.7.4
Qardio / Heart Health Android Mobile Applicationaffected 2.5.1
Qardio / QardioARMaffected All versions

CWE

CWE-248

References

ADP enrichment

CISA-ADP

No English description supplied.

No references supplied.