Description (CNA)
The IntelliSpace portal application utilizes .NET Remoting for its functionality. The vulnerability arises from the exploitation of port 755 through the deserialization vulnerability. After analyzing the configuration files, we observed that the server had set the TypeFilterLevel to Full which is dangerous as it can potentially lead to remote code execution using deserialization. This issue affects IntelliSpace Portal: 12 and prior.
Affected (CNA)
| Vendor / product | Versions |
|---|---|
| Philips / IntelliSpace Portal | affected 12 and prior |
CWE
CWE-502
References
- https://www.cve.org/CVERecord?id=CVE-2025-3425
- https://www.philips.com/a-w/security/security-advisories.html#security_advisories
ADP enrichment
No English description supplied.
No references supplied.