Description (CNA)
Medtronic MyCareLink Patient Monitor has a built-in user account with an empty password, which allows an attacker with physical access to log in with no password and access modify system functionality. This issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025
Affected (CNA)
| Vendor / product | Versions |
|---|---|
| Medtronic / MyCareLink Patient Monitor 24950 | affected 0 before June 25, 2025 |
| Medtronic / MyCareLink Patient Monitor 24952 | affected 0 before June 25, 2025 |
CWE
CWE-258
References
- https://www.medtronic.com/en-us/e/product-security/security-bulletins/mycarelink-patient-monitor-vulnerabilities.html
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-205-01
ADP enrichment
No English description supplied.