Description (CNA)
An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. They run an SSH server accessible over the default port 22. The root account has a weak default password of ionadmin, and a password change policy for the root account is not enforced. Thus, an attacker with network connectivity can achieve root code execution. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.
Affected (CNA)
| Vendor / product | Versions |
|---|---|
| n/a / n/a | affected n/a |
CWE
No CWE identifiers supplied.
References
- https://tools.thermofisher.cn/content/sfs/brochures/One_Touch_2_Spec_Sheet.pdf
- https://assets.thermofisher.com/TFS-Assets/LSG/manuals/MAN0014388_IonOneTouch2Sys_UG.pdf
- https://documents.thermofisher.com/TFS-Assets/CORP/Product-Guides/Ion_OneTouch_2_and_Torrent_Suite_Software.pdf