← CVE Intelligence

CVE-2026-17583

Thermo Fisher Applied Biosystems Genetic Analyzers Missing Support for Integrity Check PUBLISHED

Description (CNA)

The affected Thermo Fisher Applied Biosystems Genetic Analyzers are vulnerable because .fsa/.hid output files can be edited. An attacker could tamper with these files, altering DNA data and resulting in inaccurate DNA test outcomes.

Affected (CNA)

Vendor / productVersions
Thermo Fisher / Applied Biosystems 3500/3500xL Series Data Collection Softwareaffected 0 through 4.0.2
Thermo Fisher / Applied Biosystems 3730/3730xL Series Data Collection Softwareaffected 0 through 5.0.2
Thermo Fisher / Applied Biosystems SeqStudio Genetic Analyzer Data Collection Softwareaffected 0 through 1.2.5
Thermo Fisher / Applied Biosystems SeqStudio Flex Series Instrument Softwareaffected 0 through 1.2.0
Thermo Fisher / Applied Biosystems GeneMapper ID-X Softwareaffected 0 through 1.7.3
Thermo Fisher / Applied Biosystems 3130 Series Data Collection Softwareaffected 0 through 4.1
Thermo Fisher / ABI PRISM 3100/3100-Avant Data Collection Softwareaffected 0 through 2.0
Thermo Fisher / ABI PRISM 310 Data Collection Softwareaffected 0 through 3.1

CWE

CWE-353

References

ADP enrichment

CISA-ADP

No English description supplied.