Description (CNA)
The affected Thermo Fisher Applied Biosystems Genetic Analyzers are vulnerable because .fsa/.hid output files can be edited. An attacker could tamper with these files, altering DNA data and resulting in inaccurate DNA test outcomes.
Affected (CNA)
| Vendor / product | Versions |
|---|---|
| Thermo Fisher / Applied Biosystems 3500/3500xL Series Data Collection Software | affected 0 through 4.0.2 |
| Thermo Fisher / Applied Biosystems 3730/3730xL Series Data Collection Software | affected 0 through 5.0.2 |
| Thermo Fisher / Applied Biosystems SeqStudio Genetic Analyzer Data Collection Software | affected 0 through 1.2.5 |
| Thermo Fisher / Applied Biosystems SeqStudio Flex Series Instrument Software | affected 0 through 1.2.0 |
| Thermo Fisher / Applied Biosystems GeneMapper ID-X Software | affected 0 through 1.7.3 |
| Thermo Fisher / Applied Biosystems 3130 Series Data Collection Software | affected 0 through 4.1 |
| Thermo Fisher / ABI PRISM 3100/3100-Avant Data Collection Software | affected 0 through 2.0 |
| Thermo Fisher / ABI PRISM 310 Data Collection Software | affected 0 through 3.1 |
CWE
CWE-353
References
- https://documents.thermofisher.com/TFS-Assets/CORP/Product-Guides/fsa_hid_bulletin.pdf
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-216-01
- https://www.cve.org/CVERecord?id=CVE-2026-17583
ADP enrichment
No English description supplied.
No references supplied.